A manual resuscitator is not a single universal bag for every patient. The bag, valve, masks, oxygen reservoir and tubing need to form a documented, compatible set for the intended patient group. Buyers should define that set before comparing prices.
This guide helps facilities in Uganda and South Sudan prepare an equipment specification. It does not teach resuscitation. Clinical protocols, device selection and use must be approved by qualified clinical leadership, and operation requires appropriately trained staff.
Specify the patient group first
State whether the quotation is for adult, paediatric or neonatal use and whether more than one complete set is required. Ask the manufacturer to identify the resuscitator size and the mask sizes supplied for each intended group. Descriptions such as “all sizes” are not enough unless every component is listed.
Exper's manual resuscitator listing describes a hand-operated self-inflating unit with mask, valve, reservoir bag and oxygen tubing, with size options subject to confirmation. Use that page to begin an enquiry, then require the exact supplied configuration in writing.
Request the complete breathing system
- Self-inflating bag and patient valve.
- Correctly sized masks for the intended patients.
- Oxygen inlet, oxygen tubing and reservoir arrangement.
- Any pressure-limiting or pressure-relief feature specified for the model.
- Storage case, operating instructions and reprocessing instructions.
- Replacement masks, valves, reservoirs and other model-specific parts.
Ask the supplier to identify which items are included, optional or single-use. Parts from another model should not be treated as compatible because they look similar. Request manufacturer names, catalogue numbers and any required connectors.
Compare documented performance and safeguards
Ask for bag volume, delivered-volume information, pressure characteristics and the manufacturer's stated operating conditions. Have the clinical and biomedical teams review them for the intended service. Do not select settings or clinical technique from a procurement article.
The WHO technical specifications for neonatal resuscitation devices were developed to support selection, procurement, use and reprocessing, especially in lower-resource settings. They show why the correct patient group, device components and reprocessing pathway belong in the purchasing specification.
Decide whether the set is reusable or single-use
Request a clear statement for every component. For reusable equipment, obtain instructions that cover disassembly, cleaning, disinfection or sterilization, inspection, drying and reassembly. The facility's infection-prevention team must confirm that its equipment and approved products can follow those instructions.
For single-use components, plan quantities, storage and safe disposal. Do not reprocess a component labelled single-use. For reusable sets, ask how many processing cycles are permitted and which signs of wear require replacement.
Plan oxygen connection and readiness checks
Tell the supplier which oxygen source and connectors the facility uses. Ask what tubing, reservoir and adapter are required. The clinical and engineering teams should verify the complete connection and the manufacturer's readiness check before the device is placed in service.
At delivery, confirm the model, patient size and component list against the quotation. Inspect seals, valves, masks, tubing and storage case for damage. Record the agreed functional check, manuals, lot or serial identity where applicable, warranty and replacement-part information.
Include training and replacement planning
Ask what device-specific orientation is included and who provides it. Training scope should cover assembly, readiness checks, approved reprocessing and replacement criteria; clinical competency is managed by the facility. Keep the set accessible under the facility's emergency-equipment policy and record routine checks using our maintenance-records guide.
For a comparable quotation, send the patient groups, number of complete sets, oxygen connection, reusable or single-use preference, replacement quantities and destination. Browse emergency and resuscitation equipment, then contact Exper Medical Network on +256 756 789 821 or info.brandexper@gmail.com. Current availability, price, configuration, warranty and support will be confirmed in writing.